Troubleshooting the Installed Instrument
212
Troubleshooting the Instrument
“Measurement Server
Configuration not
supported” status message
Either
A Measurement Extension (M3015A or
M3016A) is connected to a standard M3000A
Measurement Server (noninvasive measure-
ments only) and an M3046A Monitor with soft-
ware Rev.D or lower. No measurements (for
example CO
2
, 2nd Press/Temp) from the Meas-
urement Extension are available.
Exchange the Measurement Server for an
M3000A with option #C06.
Or
A Measurement Extension (M3015A or M3016A)
is connected to a standard M3001A Measurement
Server (noninvasive measurements only) and an
M3046A Monitor with software Rev.E . No
measurements (for example CO
2
, 2nd Press/Temp)
from the Measurement Extension are available.
Exchange the Measurement Server for an M3001A
with option #C06 or #C18.
“Some measurements
in MMS Extension are
not supported” prompt
message
The M3046A Monitor does not support some of
the measurements in a measurement extension.
Consider using a monitor that supports these
measurements.
“Measurement Server
not supported” status
message
Either
M3001A Measurement Server is connected to an
M3046A Monitor with Rev. D or lower soft-
ware. This conbination does not allow monitor-
ing. Use only M3000A Measurement Servers
with M3046A Monitors with Rev. D or lower
software.
Or
M3000A Measurement Server is connected to an
M3046A Monitor with Rev. E software. This
conbination does not allow monitoring. Use only
M3001A Measurement Servers with M3046A
Monitors with Rev. E software.
Message
What To Do