MSL Cable Termination
Installing the Instrument
113
Step 3
Use the Insertion Tool (M3086-43801) to position each wire on the PCB according
the following wiring schematic, where each color corresponds to a number.
NOTE
The Insertion Tool should be set to cutting mode
✂
= on. The Insertion Tool automatically
removes the insulation at the connection site, clamps the wire in the correct position, and cuts
off the excess wire.
Step 4
Use a small screwdriver to connect the two drain wires to the PCB, see the wiring
schematic above.
Summary of Contents for M3000A
Page 2: ...Printed in Germany 03 05 Sixth Edition M3046 9300F Part Number M3046 9300F 4512 610 07711 S ...
Page 8: ...vi ...
Page 134: ...Answers to the Tutorial for Installing the Instrument 118 Installing the Instrument ...
Page 268: ...Monitor Parts 252 Replacement Parts Monitor Parts 5 4 3 7 2 1 6 ...