Part 3 Using Support Functions
198
Troubleshooting the Instrument
The following table explains the contents of the Status Log. The first entry on the left side of
the Status Log corresponds to the top entry under Field Title in the table.
Field Title
Description
H, C, or N
H and C—These denote fatal errors which
have caused a Hot start (like switching the
Instrument off/on) or a Cold start (like per-
forming a System Test).
N—This denotes a non-fatal error.
Device ID Code
This number refers to a software module, and
is used to indicate the location of a fault.
n
This number is the internal software module
number.
Error Code
This number describes the type of fault identi-
fied.
Date/Time
This indicates the date/time the error occurred.