Part 2 Isolating and Solving Instrument Problems
186
Troubleshooting the Instrument
Incorrect Time
Stamps
Invalid data
stored showing
wrong times for
data.
The Instrument’s real time
clock is defective.
Power on for a few min-
utes then reset the time
and date.
Replace the System
Board.
Cannot repro-
gram the time
and date
correctly
Defective System Board.
Replace the System
Board.
Compatibility related problems
“Some measure-
ments are not
supported by the
Monitor”
prompt message
A measurement extension
(M3015A or M3016A) is
connected to an M3046A
M3 monitor.
No measurements (CO
2
,
2nd Press/Temp) from the
Measurement Extension
are available.
Use extensions only with
M3046A M4 monitors