7.3 Removing components for maintenance
NOTICE
Malfunction due to changing the connection configuration
The pumping station connections are pre-configured at the factory. If the control lines on the connec-
tor are mixed up, this causes the pumping station to malfunction or fail.
► When removing components, note their original layout for subsequent re-assembly.
► Make a note of the accessories configuration and important setting values from the DCU before
disassembly of the pumping station or components.
Customer maintenance on pumping station components may require removal from the pumping station
frame.
Connectors on the TC
Preset accessory
Acc. A1
Air cooling
Acc. B1
Venting valve
DO2
Backing pump (HiCube 80 Pro only)
Tbl. 14:
Preset accessory connections
7.3.1 Carry out preparatory work
2
1
Fig. 19:
Electrical connections
1 Mains cable
2
Main switch
Procedure
1. Switch off the pumping station at the main switch.
2. Wait until all components are at a complete standstill.
3. Disconnect the mains plug.
4. For water cooling: Shut off the cooling water feed.
7.3.2 Remove turbo pump
Factory settings
All turbopump accessory connections are preconfigured at the factory. Any confusion will
cause malfunctions or operational failure.
1. Please note the initial product configuration.
2. Note the cable connections and parameter settings, before making changes.
Required tools
● Allen key
● Calibrated torque wrench (tightening factor ≤ 1.6)
Maintenance