background image

4

2.1. Main Features

2. Understanding The Rotary Vane Pumps 

2.2. Differences Between The Pump Types

Type

Stage

Nominal

Attainable

flow rate

Final Pressure

at 50 Hz

(Total, without 
Gasballast)

UNO 120 A

one stage

128 m3/h

< 3 

10-2 mbar

DUO 120 A

two-stage

128  m3/h

< 2 

10-3 mbar

The UNO 120 A is a single stage and the DUO 120 A is a two
stage rotary vane pump.
Special features:
– Pressurised oil lubrication
– Hydraulically controlled high vacuum safety valve
– Optional: Oil filter systems, Condensate separator,

Operational monitoring systems.

Proper Use

– Rotary Vane Pumps UNO/DUO 120 A may only be used to

generate vacuum.

– Do not use the pump as a compressor.
– Do not pump corrosive or explosive gases (a special versi-

on for these operations is available on request).

– Do not operate the pumps where explosive processes are

involved. This is only permitted with the "explosive" version
of the pump.

– Using accessories not named in this manual is not permit-

ted without authorisation from PFEIFFER.

Improper use will cause any rights regarding liability and gua-
rantees to be forfeited.

Special Models

Helium tight model:

– Operating Fluid P3
– Leak rate < 1 

10-5 mbar l/s

– Additional radial shaft seal with sealing oil chamber

Korrosivgasausführung:

– Operating fluid F5
– Leak rate < 1 

10-5 mbar l/s

– Additional radial shaft seal with sealing oil chamber
– Vanes made from special materials
– Gas ballast valve in special version

Rotary Vane Vacuum Pumps UNO120 A, DUO 120 A

1

Sight glass

2

Gas ballast valve, complete; 

3 Motor, 

4

Exhaust flange, 

5

Vacuum flange, 

15 Casing
18 Baffle Plate
28 Support Stand
42 Operating fluid filler screw
43 Operating medium drain screw
45 Connection socket

1 42

15

5

4

2

3

43

18

45

28

Improper Use

The following are regarded as improper:
– Uses not covered above, and, in particular:

– Connection to pumps and units which is not permitted

according to their operating instructions.

– Connection to units which contain touchable and voltage

carrying parts.

Improper use will cause any rights regarding liability and gua-
rantees to be forfeited

Summary of Contents for DUO 120 A

Page 1: ...Rotary Vane Pumps PK 800 151 BE G 0311 UNO DUO 120 A Betriebsanleitung Operating Instructions...

Page 2: ...p And Checking The Operating Fluid 9 4 1 2 Operating Fluid F5 9 4 2 Switching The Pump ON And OFF 9 4 3 Gas Ballast Valve 10 4 3 1 Standard Model 10 4 3 2 Corrosive Gas Model 10 4 3 3 Magnetic control...

Page 3: ...conversions or modifications on the pump When returning the pump to us please note the shipping instructions in Section 7 1 1 For Your Orientation Instructions in the text Operating instructions Here...

Page 4: ...s manual is not permit ted without authorisation from PFEIFFER Improper use will cause any rights regarding liability and gua rantees to be forfeited Special Models Helium tight model Operating Fluid...

Page 5: ...with operating fluid before operating the first time included in a separate pack in the delivery consignment When transporting use only the eyebolts provided for the purpose When filled with operatin...

Page 6: ...ases on the connecting cable CAUTION CAUTION Switching parts list S3 Control switch S1 Pushbutton OFF S2 Pushbutton ON K1 Contactor F1 3 Motor fuse F5 Control fuse F8 Triggering Unit TMA 4 or TMA 4R H...

Page 7: ...nection The Rotary Vane Pumps UNO DUO 120 A have been designed so that they can be connected to a Rotation Speed and Operating Hours Monitor RCI 002 which is connected at connecting socket 45 For the...

Page 8: ...ing 132 Support washer 133 Radial shaft seal 134 Double supporting ring 135 Spacer 136 Oiler 138 O ring 168 O ring 169 Locking screw 180 Hose nozzle for inert gas inlet 169 180 122 132 133 135 126 128...

Page 9: ...in final vacuum operations In non stop operations check operating fluid daily otherwise whenever the pump is switched on 4 1 2 Operating Fluid F5 Only the synthetic operating fluid F5 may be used for...

Page 10: ...ushing gas connection It has a sliding sleeve which opens the valve when pushed towards the pump and shuts the valve when pushed away from the pump Fig 8 The dosing device is closed when the groove in...

Page 11: ...FEIFFER Service Oil leaking from the foot of the pump Radial shaft seal defective Request repair by PFEIFFER Service Pump doesn t start Ambient temperature 12 C Warm up pump Dirty pump stages Request...

Page 12: ...lowing precautions Always ensure the pump cannot be switched on when carrying out any work on the pump If necessary remove the pump from the system to carry out inspection work Only dismantle the pump...

Page 13: ...O ring 34 Cover 47 Valve spindle 48 O ring 49 O ring 50 Valve housing 51 Plate spring 52 Screw 53 O ring 54 Screw 55 Insert 56 Screw 57 O ring 58 Centering sleeve 59 Hole 60 O ring 61 Sleeve 62 Flush...

Page 14: ...dosing valve unit 64 Remove insert 70 Four borings have been provided on the front side for this operation Clean all parts Check O rings 65 and 69 for wear and replace if necessary Lightly oil O rings...

Page 15: ...time is kept to a minimum Have the pump repaired on the spot by our PFEIFFER Service Engineers Return the individual components to the manufacturer for repairs Replace individual components with a new...

Page 16: ...3360 2150 Noise development without gas ballast dB A 60 60 Operating fluid amount l 17 13 Max admissible operating temperature 1 C 90 90 Nominal rotation speed of pump at 50 Hz 1 min 965 965 at 60 Hz...

Page 17: ...comprehensive range of accessories can be found in section B of the Pfeiffer catalogue order no PA 800 066 PE to extend the application range of the pumps and protect the pump from wear when used unde...

Page 18: ...he repair and or service of vacuum components will only be carried out if a correctly completed declaration has been submitted Non completion will result in delay The manufacturer could refuse to acce...

Page 19: ...t the product specified below is in accordance with the provision of EU Machinery Directive 98 37 EEC Annex II A EU Electromagentic Compatibility Directive 89 336 EEC and EU Low Voltage Directive 73 2...

Page 20: ...81 3 32 259 011 Email pfeiffer hakuto co jp Korea Pfeiffer Vacuum Korea Ltd 703 Ho 853 1 Hankuk Mulru Dongchonri Suzi Eup 449 843 Yong in City Kyungkido Phone 82 31 266 0741 Fax 82 31 266 0747 Email s...

Reviews: