Rotary Vane Vacuum Pumps UNO120 A, DUO 120 A

1

Sight glass

2

Gas ballast valve, complete; 

3 Motor, 

4

Exhaust flange, 

5

Vacuum flange, 

15 Casing
28 Support Stand

1 42 44

15

5

4

2

3

43

4.1. Filling Up And Checking The Operating

Fluid

Standard version pumps are tested with opera-
ting fluid P3. Guarantees relating to attainment
of final pressures and trouble free functioning
apply only providing this operating fluid is used.

The delivery consignment contains sufficient operating fluid
for one filling.

Pumps for special applications (e.g. for pumping corrosive
gases) can be operated with special operating fluids such as
F5 and other PFEIFFER operating fluids. The use of such
operating fluids requires prior authorisation from the
manufacturer.
P3 must be disposed of in accordance with local waste dispo-
sal regulations.

Remove operating fluid filler screw 42.

Fill up with operating fluid included in the consignment
(only for P3); refer to rating plate for correct quantity.

Replace operating fluid filler screw 42; be careful with the
O-ring.

Check operating fluid level only when the pump is warm
and running and vacuum connection 5 and gas ballast val-
ve are 

closed.

The level of operating fluid must always be between the
top and the middle of the sight glass 1.

CAUTION

Topping up operating fluid is possible when the pump is
running and in final vacuum operations.

In non-stop operations check operating fluid daily,
otherwise whenever the pump is switched on.

4.1.2. Operating Fluid F5

Only the synthetic operating fluid F5 may be used for corrosi-
ve gas model pumps.

The use of other operating fluid must be cleared by the manu-
facturer.

When working with synthetic operating media,
the manufacturer’s directions must be
observed.
If ‘’F5’’ is heated above 300°C, toxis vapours
which can harm the airways are given off.
Do not allow ‘’F5’’ to come into contact with

tobacco products (danger of poisoning when ignited). The
precautions necessary in the handling of chemicals must be
observed.

Please note:

If there is a noticeable loss of operating fluid, the radial shaft
seals should be checked.
If operating fluid is leaking out underneath the casing 15 from
the opening between the support stand 28 and motor 3, the
radial shaft seal must be replaced. Please contact your loca
Pfeiffer-Service (see Section 7.).

4.2. Switching The Pump ON And OFF

When operating the pump gets hot.

The pump can be switched ON and OFF in any pressure
range. On switching off, the built-in high vacuum safety valve
6 to the vacuum plant closes automatically and vents the
pump.
Minimum start-up temperature, in compliance with German
Industrial Standard DIN 28 426: + 12 °C.
On switching OFF (intake pressure < 750 mbar) the built-in
high vacuum safety valve to the vacuum chamber closes
automatically and the pump is vented.

CAUTION

WARNING

4. Operations 

42 Operating fluid filler 

screw

43 Operating medium drain 

screw

28

Summary of Contents for DUO 120 A

Page 1: ...Rotary Vane Pumps PK 800 151 BE G 0311 UNO DUO 120 A Betriebsanleitung Operating Instructions...

Page 2: ...p And Checking The Operating Fluid 9 4 1 2 Operating Fluid F5 9 4 2 Switching The Pump ON And OFF 9 4 3 Gas Ballast Valve 10 4 3 1 Standard Model 10 4 3 2 Corrosive Gas Model 10 4 3 3 Magnetic control...

Page 3: ...conversions or modifications on the pump When returning the pump to us please note the shipping instructions in Section 7 1 1 For Your Orientation Instructions in the text Operating instructions Here...

Page 4: ...s manual is not permit ted without authorisation from PFEIFFER Improper use will cause any rights regarding liability and gua rantees to be forfeited Special Models Helium tight model Operating Fluid...

Page 5: ...with operating fluid before operating the first time included in a separate pack in the delivery consignment When transporting use only the eyebolts provided for the purpose When filled with operatin...

Page 6: ...ases on the connecting cable CAUTION CAUTION Switching parts list S3 Control switch S1 Pushbutton OFF S2 Pushbutton ON K1 Contactor F1 3 Motor fuse F5 Control fuse F8 Triggering Unit TMA 4 or TMA 4R H...

Page 7: ...nection The Rotary Vane Pumps UNO DUO 120 A have been designed so that they can be connected to a Rotation Speed and Operating Hours Monitor RCI 002 which is connected at connecting socket 45 For the...

Page 8: ...ing 132 Support washer 133 Radial shaft seal 134 Double supporting ring 135 Spacer 136 Oiler 138 O ring 168 O ring 169 Locking screw 180 Hose nozzle for inert gas inlet 169 180 122 132 133 135 126 128...

Page 9: ...in final vacuum operations In non stop operations check operating fluid daily otherwise whenever the pump is switched on 4 1 2 Operating Fluid F5 Only the synthetic operating fluid F5 may be used for...

Page 10: ...ushing gas connection It has a sliding sleeve which opens the valve when pushed towards the pump and shuts the valve when pushed away from the pump Fig 8 The dosing device is closed when the groove in...

Page 11: ...FEIFFER Service Oil leaking from the foot of the pump Radial shaft seal defective Request repair by PFEIFFER Service Pump doesn t start Ambient temperature 12 C Warm up pump Dirty pump stages Request...

Page 12: ...lowing precautions Always ensure the pump cannot be switched on when carrying out any work on the pump If necessary remove the pump from the system to carry out inspection work Only dismantle the pump...

Page 13: ...O ring 34 Cover 47 Valve spindle 48 O ring 49 O ring 50 Valve housing 51 Plate spring 52 Screw 53 O ring 54 Screw 55 Insert 56 Screw 57 O ring 58 Centering sleeve 59 Hole 60 O ring 61 Sleeve 62 Flush...

Page 14: ...dosing valve unit 64 Remove insert 70 Four borings have been provided on the front side for this operation Clean all parts Check O rings 65 and 69 for wear and replace if necessary Lightly oil O rings...

Page 15: ...time is kept to a minimum Have the pump repaired on the spot by our PFEIFFER Service Engineers Return the individual components to the manufacturer for repairs Replace individual components with a new...

Page 16: ...3360 2150 Noise development without gas ballast dB A 60 60 Operating fluid amount l 17 13 Max admissible operating temperature 1 C 90 90 Nominal rotation speed of pump at 50 Hz 1 min 965 965 at 60 Hz...

Page 17: ...comprehensive range of accessories can be found in section B of the Pfeiffer catalogue order no PA 800 066 PE to extend the application range of the pumps and protect the pump from wear when used unde...

Page 18: ...he repair and or service of vacuum components will only be carried out if a correctly completed declaration has been submitted Non completion will result in delay The manufacturer could refuse to acce...

Page 19: ...t the product specified below is in accordance with the provision of EU Machinery Directive 98 37 EEC Annex II A EU Electromagentic Compatibility Directive 89 336 EEC and EU Low Voltage Directive 73 2...

Page 20: ...81 3 32 259 011 Email pfeiffer hakuto co jp Korea Pfeiffer Vacuum Korea Ltd 703 Ho 853 1 Hankuk Mulru Dongchonri Suzi Eup 449 843 Yong in City Kyungkido Phone 82 31 266 0741 Fax 82 31 266 0747 Email s...

Reviews: