Unpacking and Installation . 29
2.
Carefully remove the instrument from the shipping container, but not from the bag in
which it was shipped.
Retain the original packaging in case you need to ship the instrument to a new location.
CAUTION
ATTENTION
The spectrometer should be allowed to reach the temperature of its
surroundings before it is removed from the protective bag to prevent
condensation forming on the CaF
2
windows. This means leaving it for at
least 3 hours after removal from the shipping container.
Le spectromètre doit être autorisé à atteindre la température de son
environnement avant d'être retiré du sac de protection afin d'empêcher
la formation de condensation sur les fenêtres CaF2. Cela signifie le laisser
pendant au moins 3 heures après le retrait du conteneur d'expédition.
3.
When the spectrometer has been allowed to warm to the temperature of its
surroundings, remove it from the bag and place it on the bench where it is to be used.
Ensure that you can reach the rear of the spectrometer to enable you to connect the
cables.
NOTE:
Do not connect your instrument to the PC using the USB cable provided until you
have run the software installation wizard.
Any accessories will be shipped in separate boxes.
Summary of Contents for SPECTRUM TWO N
Page 1: ...SPECTRUM TWO N User s Guide MOLECULAR SPECTROSCOPY...
Page 5: ...Introduction...
Page 10: ...10 Spectrum Two N User s Guide...
Page 11: ...Warnings and Safety Information...
Page 19: ...An Overview of the Spectrum Two N...
Page 25: ...Unpacking and Installation...
Page 47: ...Getting Started with Spectrum Software...
Page 93: ...Getting Started with Spectrum Software 93...
Page 94: ......
Page 95: ...Maintenance...
Page 110: ...110 Spectrum Two N User s Guide 4 Follow the instructions on screen...
Page 116: ......
Page 117: ...Appendices...
Page 125: ...Index 125...