62 . Spectrum Two N User's Guide
Compare
The Compare process estimates similarity between spectra by comparison with a reference
spectrum or spectra, and is used to confirm the identity of a sample. The results are
displayed on the Compare tab in the Viewing Area (Figure 40).
Figure 42 Compare tab
In this example, the spectrum has been confirmed to be the same as the reference
spectrum, where the Correlation was defined at 0.98. The Correlation result is greater than
0.99, where 1.00 would mean that the spectra were identical.
NOTE:
For an example of a Report generated showing the Compare results, refer to
on page 75.
The parameters used in the Compare process are defined on the Setup Compare tabs.
Select
Compare
from the Setup menu to display the Setup Compare tabs.
On the Setup Compare References tab (Figure 41) you can load a single spectrum file or a
folder of spectra. Then set the Correlation and/or Discrimination thresholds to define the
Pass/Fail result.
Correlation expresses the degree of similarity between the spectra, where 1 is identical, and
0 is wholly unrelated. Discrimination ensures that the closest match is sufficiently separated
from the next closest match. Its value is the difference in the Correlation values of the first
two results on the Compare tab.
Only files or folders with
Include
enabled will be used in the Compare.
Figure 43 Setup Compare References tab
Summary of Contents for SPECTRUM TWO N
Page 1: ...SPECTRUM TWO N User s Guide MOLECULAR SPECTROSCOPY...
Page 5: ...Introduction...
Page 10: ...10 Spectrum Two N User s Guide...
Page 11: ...Warnings and Safety Information...
Page 19: ...An Overview of the Spectrum Two N...
Page 25: ...Unpacking and Installation...
Page 47: ...Getting Started with Spectrum Software...
Page 93: ...Getting Started with Spectrum Software 93...
Page 94: ......
Page 95: ...Maintenance...
Page 110: ...110 Spectrum Two N User s Guide 4 Follow the instructions on screen...
Page 116: ......
Page 117: ...Appendices...
Page 125: ...Index 125...