![PerkinElmer RamanFlex 400 Getting Started Manual Download Page 80](http://html1.mh-extra.com/html/perkinelmer/ramanflex-400/ramanflex-400_getting-started-manual_1530232080.webp)
80 . RamanFlex 400 Series Getting Started Guide
SOP-8: Wavelength Calibration Verification of the
RamanFlex 400
1.
Ensure the instrument is switched on by following SOP-1.
2.
Open the Scan and Instrument Setup dialog and then click Spectral Preview.
NOTE: If you have a RamanFlex 400F with two fiber optic probes, ensure that the probe not
in use is covered.
3.
Focus on a sample of polystyrene (L1181228).
4.
When the polystyrene sample is optimally focused, close the Scan and Instrument Setup
dialog by clicking Exit.
5.
Select Validate
and then Instrument from the Instrument menu in Spectrum.
The Validation / Calibration dialog is displayed.
6.
Select the Verification tab.
7.
Select Wavelength Calibration Verification, and then click OK.
The wavelength calibration verification routine begins. This will take a few minutes.
After completing the wavelength calibration verification routine, an audit log is
generated.
8.
Remove the polystyrene sample and store safely.
The instrument is now ready for use.
9.
Ensure that the probe is covered.
In the unlikely event that the wavelength calibration verification should fail, follow Figure 18.
Further Details on the wavelength calibration verification are provided as a spectral (*.sp)
file, which is saved in
C:\pel_data\config\Raman\Calibration Files\Logs
folder.
Summary of Contents for RamanFlex 400
Page 1: ...RamanFlex 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 10: ...10 RamanFlex 400 Series Getting Started Guide ...
Page 11: ...Safety Practices ...
Page 27: ...Raman History Theory and Instrumentation ...
Page 34: ...34 RamanFlex 400 Series Getting Started Guide ...
Page 35: ...Preparing your Facility ...
Page 40: ...40 RamanFlex 400 Series Getting Started Guide ...
Page 41: ...Installing your RamanFlex 400 ...
Page 61: ...Introduction to the RamanFlex 400 ...
Page 68: ...68 RamanFlex 400 Series Getting Started Guide ...
Page 69: ...Using the RamanFlex 400 ...