Installing your RamanFlex 400 . 59
31.
Click Finish.
The Raman Instrument Installer closes.
When the installation is complete, the InstallShield Wizard Completed page is displayed.
32.
Click Finish.
The Raman Instrument Installer Information dialog is displayed.
33.
Click
Yes
to restart your PC.
NOTE: You should not allow the PC to enter Standby mode. On the Screen Saver page of
the Display Properties dialog, click Power and then set System standby and Turn
off hard disks to Never.
This completes the installation of the Spectrum software.
If appropriate, you should now install the triggered fiber optic probe drivers, as described in
the
Raman Triggered Fiber Optic Probe
leaflet (L1321887), and then connect the fiber optic
cables from the triggered fiber optic probe to the RamanFlex 400, as described for a
Attaching and Removing the Fiber Optic Probe
on page 64.
If are using a RamanFlex 400 in conjunction with a RamanMicro 300 you should refer to the
RamanMicro 300 Getting Started Guide
(L1321862) for details of how to install the camera
drivers.
Summary of Contents for RamanFlex 400
Page 1: ...RamanFlex 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 10: ...10 RamanFlex 400 Series Getting Started Guide ...
Page 11: ...Safety Practices ...
Page 27: ...Raman History Theory and Instrumentation ...
Page 34: ...34 RamanFlex 400 Series Getting Started Guide ...
Page 35: ...Preparing your Facility ...
Page 40: ...40 RamanFlex 400 Series Getting Started Guide ...
Page 41: ...Installing your RamanFlex 400 ...
Page 61: ...Introduction to the RamanFlex 400 ...
Page 68: ...68 RamanFlex 400 Series Getting Started Guide ...
Page 69: ...Using the RamanFlex 400 ...