16 . RamanFlex 400 Series Getting Started Guide
Electrical Safety
The spectrometer module comes assembled within a protective housing to prevent exposure
to electrical hazards. There are also a number of exterior service panels that you should
remove only if you are cleaning the air filters. For more information see
Cleaning the Air
Filters
on page 87.
Electrical Protection
Insulation: Class I rating for external circuits. Only connect equipment that meets the
requirements of IEC 61010-1, IEC 60950 or equivalent standards.
Installation Category: The instruments are able to withstand transient overvoltages typically
present on the MAINS supply. The normal level of transient overvoltages is impulse
withstand (overvoltage) category II of IEC 60364-4-443.
Pollution Degree 2: Normally only non-conductive POLLUTION occurs. Occasionally, however,
a temporary conductivity caused by condensation must be expected.