38 . RamanFlex 400 Series Getting Started Guide
Finding a Suitable Location for the Instrument
The following conditions should be observed:
•
There must be a minimum of 3 power points (1 for the spectrometer, 2 for the PC). If
your system has a deep depletion CCD you will need 2 additional power points (1 for
the CCD PSU and 1 for the recirculating water cooler).
•
The workbench must be flat and level.
•
The workbench where the spectrometer and PC are to be located should be at a
suitable height to allow the user to work comfortably.
•
Use conducting floor coverings to minimize static discharge.
•
Floor vibrations or noise from heavy manufacturing equipment can affect the
performance of the instrument.
•
Intense magnetic fields should be at least 5.5 meters (18 feet) away from the
spectrometer.
•
Never place the instrument near a window that gets direct sunlight.
We recommend that you install the instrument in a room with no windows which meets the
appropriate safety requirements for a Class 3B laser product.
Ensure all users understand laser safety issues before locating and using the instrument.
Consult your laser safety officer and perform a risk assessment before installing the
equipment.
Summary of Contents for RamanFlex 400
Page 1: ...RamanFlex 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 10: ...10 RamanFlex 400 Series Getting Started Guide ...
Page 11: ...Safety Practices ...
Page 27: ...Raman History Theory and Instrumentation ...
Page 34: ...34 RamanFlex 400 Series Getting Started Guide ...
Page 35: ...Preparing your Facility ...
Page 40: ...40 RamanFlex 400 Series Getting Started Guide ...
Page 41: ...Installing your RamanFlex 400 ...
Page 61: ...Introduction to the RamanFlex 400 ...
Page 68: ...68 RamanFlex 400 Series Getting Started Guide ...
Page 69: ...Using the RamanFlex 400 ...