DISPLAY INFO
PAtH02
rEV. 04
When the unit is switched on, a display test is performed; the software ID and the
relevant version are then displayed. Such codes should be communicated in case of
support request.
When the set-up procedure is not under way, the display shows the measured weight in kg. The following
messages appear in certain conditions:
ERROR REPORTING
The following error codes may appear on the display in the operating mode.
Fixed message
O-L
No load cell signal, or signal falls outside the mV/V measurement field.
- - - -
- -
Overload. The weight applied to the load cells exceeds the maximum capacity of the
weighing system by over 9 divisions.
- - - - - -
Underload. This message appears when the measured weight is negative below the
displayable value.
Flashing message, alternating with the measured weight
no-CAL
Weight calibration not performed, flashing message (alternating with the measured
weight).
Low.Uin
Power supply too low in case of operation with power unit, flashing message (alternating
with the measured weight).
E.FibEr
Communication error with the S318 optical fiber interface board. This error is only
displayed in case of communication timeout and when the interface is enabled (ana-
logical output option enabled, COM3 protocol enabled, or COM4 protocol enabled).
No CoN
Fieldbus network disconnected.
E-F.buS
Fieldbus interface connection error.