
Page 30
C3Prot.
COM3 PROTOCOL
Defines the usage mode of the RS232 serial port on an external S318 optical fiber interface:
Selectable values:
None:
Serial communication disabled
Contin:
Continuous transmission of the weight string. Can be used, among other things, to drive a
weight repeater. See the relevant section for details.
Slave:
ASCII protocol. See the relevant section for details.
Modbus:
MODBUS RTU protocol. See the relevant section for details.
on deM:
A weight string is transmitted when the operator presses the relevant front key or via Input
2. The command is only executed if weight is stable. Weight should vary by at least 20 divisions
between two subsequent transmissions.
Default: Slave
C3bAud.
COM3 BAUD RATE
Defines the baud rate of the RS232 serial port on an external S318 optical fiber interface.
The value should be set to the same value as the PC/PLC or the remote display.
Selectable values:
1200; 2400; 4800; 9600; 19200; 38400; 57600; 115200
Default: 9600
C3FrAN
COM3 PROTOCOL
Frame type. In case of SLAVE or MODBUS protocol, the 7-bit (E-7-1 and O-7-1) data format cannot be
selected:
Selectable values:
n-8-1; n-8-2; E-7-2; E-8-1; o-7-2; o-8-1
Default: n-8-1
Summary of Contents for MC 315 Series
Page 2: ...Page II ...
Page 61: ...Page 59 ...