Page 31
C4Prot.
COM4 PROTOCOL
Defines the usage mode of the RS485 serial port on an external S318 optical fiber interface:
Selectable values:
None:
Serial communication disabled
Contin:
Continuous transmission of the weight string. Can be used, among other things, to drive a
weight repeater. See the relevant section for details.
Slave:
ASCII protocol. See the relevant section for details.
Modbus:
MODBUS RTU protocol. See the relevant section for details.
on deM:
A weight string is transmitted when the operator presses the relevant front key or via Input
2. The command is only accepted if weight is stable. Weight should vary by at least 20 divisions
between two subsequent transmissions.
Default: Slave
C4bAud.
COM4 BAUD RATE
Defines the baud rate of the RS485 serial port on an external S318 optical fiber interface.
The value should be set to the same value as the PC/PLC or the remote display.
Selectable values:
1200; 2400; 4800; 9600; 19200; 38400; 57600; 115200
Default: 9600
C4ForN
COM3 PROTOCOL
Frame type. In case of SLAVE or MODBUS protocol, the 7-bit (E-7-1 and O-7-1) data format cannot be
selected:
Selectable values:
n-8-1; n-8-2; E-7-2; E-8-1; o-7-2; o-8-1
Default: n-8-1
AddrES
COMMUNICATION ADDRESS
Communication address of the unit:
Values: 1 to 32 Default: 1