Ⅳ.
IN CASE OF TROUBLE
81
2. Product error indication with LED lighting
LED lighting list
If a failure occurs, the LED indicators on the side-rail integrated panel, the nurse control panel (accessory),
or the handheld controller may flash to indicate an error condition. Check the list below.
<Bed operation>
Problem
Power LED
flashing rate
Cause
Corrective action
The LED indicators
flash when a button on
the side-rail integrated
panel, patient control
panel, handheld
controller, or nurse
control panel is
pressed.
Flashing
(0.2-second
interval)
The load on the bed
exceeded the safe
working load.
Check that the load on the bed is
within the safe working load.
Operation of the bed is
restricted by an
obstacle.
Make sure that operation of the bed is
not restricted by an obstacle.
The linear actuator in
the bed is causing an
error.
(The control unit cannot
detect the height of the
bed.)
This error can be solved by setting the
bed height to the minimum height.
Press the height adjustment
and
buttons simultaneously at the
same time to make the buzzer sound.
Press the buttons until the buzzer
sound stops to release the error status
and resume normal operation. If the
error is not resolved, contact
Paramount Bed (see p.100) or your
distributor.
The linear actuator on
the bed is
disconnected.
Contact Paramount Bed (see p.100)
or your distributor.
The cable is broken.
Contact Paramount Bed (see p.100)
or your distributor.
Pressing any button
other than the bed
operation buttons for 30
seconds or longer.
Release all pressed buttons and wait
for 10 seconds for the error to be
resolved.
Bed does not work.
LED indicator
does not
flash
Two or more bed
operation buttons are
pressed
simultaneously.
Release all the pressed buttons.
Three or more optional
buttons are pressed
simultaneously.
Release all the pressed buttons.
The bed operation
exceeded the
maximum continuous
operating time.
Disconnect the power plug from the
outlet and leave the bed for 10
minutes. There is a possibility that a
circuit to protect the electrical
components against overheating was
activated. This is not a failure.
If the problem persists after taking the corrective action above, contact Paramount
Bed (see p.100) or your distributor.