52
Electric CPR
Hold down the
button (8) for at least 2 seconds or press it twice quickly to operate the bed so the
head/knee sections and inclination angles b
ecome 0˚ (horizontal).
* The bed is lowered to the minimum height.
* Releasing the button stops the operation at its position.
Patient lock-out
Press the
button (9) to disable the operation of the patient control panel and handheld controller
(accessory).
Press the
button (9) again to cancel the lock-out status.
* The LED indicator is on when locked.
Overall lock-out
Press the
button (10) to disable all bed operations except for electric CPR.
Press the
button (10) again to cancel the lock-out status.
* The LED indicator is on when locked.
* Even in the patient lock-out state, the overall lock-out state can be set by pressing the button. When the
overall lock-out state is canceled, the patient lock-out state is also canceled.
Pressing two or more buttons simultaneously stops the adjustment. If you stop
the adjustment by pressing two or more buttons, release your finger from all
buttons and then perform the adjustment again.
When the TR/RTR button is selected with the head end raised and
is pressed,
the movement stops when the bed becomes flat. To incline the bed further, press
again.
When the TR/RTR button is selected with the head end lowered and is
pressed, the movement stops when the bed becomes flat. To incline the bed
further, press
again.
In order to protect electrical parts from overheating, the buttons cannot be
operated if the bed is used continuously. In that case, disconnect the power
cable, and leave the bed for 10 minutes.
If the Electric CPR button is pressed, the head section flattens first, and then the
bed lowers to the minimum height and the knee section flattens.