82
�
.
REF ERENC E INFORM AT ION
1. Regular inspection and daily maintenance
Regular inspection
To ensure safety, inspect the product regularly, at least once a year. The product wears out and
deteriorates differently depending on the operating frequency and environment. Inspect for loose parts
and damage and the operation of the moving parts regularly. If you notice an irregularity or want a
detailed inspection, contact Paramount Bed (see p.100) or your distributor.
Part
Operation
Inspection item
Bed
Head up operation
* See p.42
Proper operation can be achieved.
There must not be any strange noises.
Knee up operation
* See p.42
Proper operation can be achieved.
There must not be any strange noises.
Height adjustment
operation
* See p.42
Proper operation can be achieved.
There must not be any strange noises or significant
rattling.
Bed operation buttons
* See p.49 to 53
All bed operation buttons must work.
CPR
* See p.56
The head section must be lowered with a CPR lever.
Miscellaneous
There must not be any deformation, damage, or
detachment of roller bumpers or resin parts.
There must not be any peeling of labels, and the text
must be legible.
There must not be any scratches or deformation at
the mattress deck, frame, or any other parts.
Board
Attachment state
* See p.24
The board stopper must be locked properly.
There must not be any significant rattling.
Caster
Traveling
There must not be any strange noises when the bed
is traveling.
There must not be any abnormal vibrations when the
bed is traveling.
Caster locking operation
* See p.29
Proper operation can be achieved.
The operation must not be overly difficult.
The bed must not move when casters are locked.
The bed must move smoothly when the lock is
released.
When in the steering position, swiveling of the
steering caster must be locked.
Side-rail
Side-rail operation
* See p.25
Proper operation can be achieved.
There must not be any strange noises or significant
rattling.
The operation must not be overly difficult.
Electrical components Visual check of electrical
components
Cables must not get tangled in the bed.
The plug and the cable must have no scratches or
damage.
The side-rail integrated panel, patient control panel,
handheld controller (standard or accessory), and
nurse control panel (accessory) must have no
scratches or damage.
Brightness of night-light does not decrease.