Ⅰ.
BEFORE USING THE BED
19
Check around the bed before operating the bed.
Bed operation may damage objects around the bed, or may cause the bed to be damaged or
deformed.
Make sure that the board stoppers of the headboard and footboard
are locked.
If the board is not attached properly, the board
may suddenly disengage when the board is
pushed or pulled to move the bed or the patient
grabs the board to support their body, and the
nurse or patient may lose their balance and fall
down, resulting in injury. (See p.24)
Do not move the bed while holding the side-rails.
This applies excessive force to the side-rails, resulting in damage or deformation of the side-rails.
Avoid moving the bed over uneven floors.
Do not place the bed on the inclination which angle is 6 degree or
more.
The bed may tip over, resulting in an injury.
Avoid moving the bed over uneven floors.
This may cause damage to the bed, or
deformation or failure of the bed. If you must move
the bed, move it as slowly as possible.
Attach a handheld controller (standard or accessory) or nurse
control panel (accessory) suitable for the bed.
If a handheld controller or nurse control panel not suitable for the bed is attached by mistake, the
bed may malfunction.
A suitable handheld controller or nurse control panel may be different among beds of the same
series due to specification changes. Before attaching a handheld controller or nurse control panel
to the bed, check the model number and shape of the handheld controller or nurse control panel to
make sure that they are the same as those of the handheld controller or nurse control panel
suitable for the bed. (See p.59 and p.61)