<Checking the battery status>
The battery status is displayed as shown in the table below.
Power status
Battery status indicator LED
Battery status
The power plug is not
inserted into the outlet
or power failure is
occurring
Lights in green
Remaining battery power is sufficient.
Flashes green
(on for 2 seconds → off for 2 seconds)
Remaining battery power is low. Since bed
operation may become impossible, stop
using and recharge the battery as soon as
possible.
* The buzzer sounds if the buttons are
pressed while the battery power is low or
insufficient.
Flashes green
(on for 0.1 seconds → off for 3.9 seconds)
Remaining battery power is insufficient.
Buttons on the side-rail integrated panel,
nurse control panel (accessory), and
handheld controller do not work. Recharge
the battery immediately.
* The buzzer sounds if the buttons are
pressed while the battery power is low or
insufficient.
The power plug is
inserted into the outlet
Flashes green
(on for 3.9 seconds → off for 0.1 seconds)
Battery is being recharged.
Lights in green
Battery is fully charged.
To extend the battery life, recharge the battery as frequently as possible before the
indicator shows that the remaining power is low or insufficient (see the table above).