CHAPTER 5 FUNCTION ITEMS
5-25
5. Set the lower limit value for sensor head 2.
Touch
again.
Set the lower limit value in the lower limit value screen
for sensor head 2.
6. Return to the MENU screen by touching
or
return to the input/output screen by touching
.
Reference
Input Range: Upper Limit Value, Lower Limit Value = ±999.9999 mm
Set values so the upper limit becomes larger than the lower limit.
Default Settings:
Sensor Head 1 Upper Limit Value = Add/subtract Calculation Upper Limit Value =
+5. 20.0000mm
Sensor Head 1 Lower Limit Value = Add/subtract Calculation Lower Limit Value =
-5.0000mm/- 20.0000mm
Sensor Head 2 Upper Limit Value = +5. 20.0000mm
Sensor Head 2 Lower Limit Value = -5.0000mm/- 20.0000mm (50mm type/85mm type)
Hysteresis Setting
Set hysteresis for upper and lower limit values to prevent chattering. Because the
resolution changes based on the workpiece measured, the average number of samples,
and so on, set operating conditions and use the P-P mode to obtain an estimated
resolution before setting hysteresis. Hysteresis can be set for each sensor head
.
1. Display the upper limit hysteresis screen from the
input/output screen.
Touch
on the MENU Screen. The input/output screen
is displayed.
Touch
on the input/output screen to open the upper
limit hysteresis screen. The current setting and a keyboard for
inputting value are displayed.
2. Set the upper limit hysteresis.
Input hysteresis for the upper limit value of sensor head 1.
Hysteresis works evenly in the upper and lower value areas
around the upper limit value.
For details, see the description "Hysteresis" below.
For keyboard input method, see page 4-2.
Input/output Screen
Upper limit hysteresis Screen
Summary of Contents for HL-C1 Series
Page 7: ...3 Warning Label Laser Emission Indicator Measurement Range Indicator ...
Page 18: ...CHAPTER 1 NAMES AND FUNCTIONS OF PARTS ...
Page 25: ...CHAPTER 1 NAMES AND FUNCTIONS OF PARTS 1 7 MEMO ...
Page 26: ...CHAPTER 2 INSTALLATION ...
Page 32: ...CHAPTER 2 INSTALLATION 2 6 z Measurement of specular reflection 50mm type 85mm type ...
Page 39: ...CHAPTER 2 INSTALLATION 2 13 MEMO ...
Page 40: ...CHAPTER 3 INPUT AND OUTPUT TERMINAL BLOCKS ...
Page 49: ...CHAPTER 3 INPUT AND OUTPUT TERMINAL BLOCKS 3 9 MEMO ...
Page 50: ...CHAPTER 4 MEASUREMENT ...
Page 54: ...CHAPTER 4 MEASUREMENT 4 4 1 2 4 6 3 5 ...
Page 60: ...CHAPTER 5 FUNCTION ITEMS ...
Page 109: ...CHAPTER 5 FUNCTION ITEMS 5 49 MEMO ...
Page 110: ...CHAPTER 6 RS 232C CONTROL ...
Page 149: ...CHAPTER 6 RS 232C CONTROL 6 39 MEMO ...
Page 150: ...CHAPTER 7 INSPECTION AND MAINTENANCE ...
Page 152: ...CHAPTER 8 TROUBLESHOOTING ...
Page 156: ...CHAPTER 9 SPECIFICATIONS ...
Page 162: ...CHAPTER 10 DIMENSIONS ...
Page 164: ...CHAPTER 10 DIMENSIONS 10 2 z 85mm diffused reflection z 85mm specular reflection ...
Page 166: ...CHAPTER 10 DIMENSIONS 10 4 z 85mm diffused reflection z 85mm specular reflection ...
Page 168: ...CHAPTER 10 DIMENSIONS 10 6 10 2 Controller Unit mm 5 5 8 1 4 M3 55 104 120 74 40 Dpth 1 5 ...