4.8.5.6 Perform AutoCal?
The
Perform AutoCal? option can activate the Auto Cal procedure without waiting for the specified Auto Cal interval.
To set Auto Cal to begin automatically, refer to page 42.
When the TMO2D performs an Auto Cal, either upon user command or at a prespecified time, it performs several
operations.
•
The Process/Cal relay will trip to the Cal position, and the Zero/Span relay will trip to the Span position. Refer to
the Relays Menu (page 31) to configure relays for calibration control.
For a 1 gas cal configuration:
•
The first line of the display reads either “AutoCal: Span Gas is ON” or “AutoCal: Offset Gas is ON” (depending on
the active mode) along with the equilibration time, which counts down to zero.
You can interrupt the Auto Cal at this point by pressing [NO]. The Display will prompt
Abort AutoCal? You have two
choices.
•
Press [NO] to resume the countdown.
•
Press [YES] to return the display to the beginning of the
Perform Auto Cal? option.
When the equilibration time reaches zero, the display will read
Measuring . . . and count down the measurement time.
At the end of the Auto Cal procedure, the Process/Cal relay will reset to the Process position and the settling time will
count down to zero. The display will then read
AutoCal in Progress and the TMO2D will return to the start of the
Perform Auto Cal? option.
For a 2-gas-cal configuration:
•
The first line of the display reads “AutoCal: Span Gas is ON” along with the equilibration time, which counts down
to zero.
You can interrupt the Auto Cal at this point by pressing [NO]. The Display will prompt
Abort AutoCal? You have two
choices.
•
Press [NO] to resume the countdown.
•
Press [YES] to return the display to the beginning of the
Perform Auto Cal? option.
When the equilibration time reaches zero, the display will read “Measuring . . .” and count down the measurement
time. At the end of the measurement, the display changes to “AutoCal: Zero Gas is ON”, along with its equilibration
time, which also counts down to zero.
At the end of the Auto Cal procedure, the Process/Cal relay will reset to the Process position and the settling time will
count down to zero. The display will then read “AutoCal in Progress” and the TMO2D will return to the start of the
Perform Auto Cal? option.
Summary of Contents for TMO2D
Page 1: ...panametrics com Process Analyzers 910 084 D August 2021 TMO2D Digital Display User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Chapter 1 Features and Capabilities 2 TMO2D User s Manual no content intended for this page...
Page 24: ...Chapter 2 Installation 16 TMO2D User s Manual no content intended for this page...
Page 26: ...Chapter 2 Installation TMO2D User s Manual 18...
Page 64: ...Chapter 4 Basic Programming 56 TMO2D User s Manual no content intended for this page...
Page 74: ...Chapter 6 Specifications 66 TMO2D User s Manual no content intended for this page...
Page 87: ...Appendix A Outline and Installation Drawings TMO2D User s Manual 79...
Page 89: ...Appendix B Menu Maps 81 TMO2D User s Manual no content intended for this page...
Page 94: ...Appendix B Menu Maps TMO2D Digital Display User s Manual 86 Figure 30 Factory Setup Menu...
Page 95: ...Appendix B Menu Maps TMO2D Digital Display User s Manual 87...
Page 99: ...76 TMO2D User s Manual no content intended for this page...
Page 100: ......