Chapter 4. Basic Programming
48
TMO2D User’s Manual
4.8.5.4 Set Settling Time?
In the Set Settling Time option, you can enter (in minutes and seconds) the settling time, the amount of time the
process gas should remain on before the TMO2D resumes taking measurements and reactivates the recorder
output. (During an autocal, the recorder output from the TMO2D has been locked to the last value it had sent directly
before the autocal began.)
.
4.8.5.5 Set Error Handling?
Use the
Set Error Handling? option to program the display and the recorder outputs to respond to any error (invalid
measurement) that occurs during Auto Cal. Errors occur if the measured drift values exceed preprogrammed limits
(as explained on page 42). To program alarm relays to respond to calibration errors, go to the Relays Menu (page 31).
Follow these steps to enable error handling, set the display screen, and select the output responses.
.
AUTO CALIBRATION MENU
Set Settling Time?
Scroll through the submenu, and press [YES] to enter the Set Settling Time option.
Settling Time:
MM.SS [XX.XX]:
Use the numeric keys to enter the minutes and seconds for settling time, and press
[YES] twice to confirm the entry. Press [NO] to proceed.(The X’s represent the
previously entered time.)
AUTO CALIBRATION MENU
Set Settling Time?
Press [NO] to proceed to the “Set Error Handling?” option.
AUTO CALIBRATION MENU
Set Error Handling?
Scroll through the submenu, and press [YES] to enter the Set Error Handling option.
Enable Error Handling
[YES] no
Use the [NO] or arrow keys to scroll to [YES], and press [YES] to confirm the
selection.
AUTOCAL ERROR EFFECTS
Set Display Response?
Press [YES] to set the Display Response.
Display AutoCal Error?
[YES] no
Use the [NO] or arrow keys to make the desired selection, and then press [YES] to
confirm the entry.
AUTOCAL ERROR EFFECTS
Set Display Response?
Press [NO] to proceed to the next item.
AUTOCAL ERROR EFFECTS
Set Output Response?
Press [YES] to set the Output Response.
Set Output Response
No Effect?
Use the [NO] or arrow keys to scroll through the choices, then press [YES] at the
desired output response.
Summary of Contents for TMO2D
Page 1: ...panametrics com Process Analyzers 910 084 D August 2021 TMO2D Digital Display User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Chapter 1 Features and Capabilities 2 TMO2D User s Manual no content intended for this page...
Page 24: ...Chapter 2 Installation 16 TMO2D User s Manual no content intended for this page...
Page 26: ...Chapter 2 Installation TMO2D User s Manual 18...
Page 64: ...Chapter 4 Basic Programming 56 TMO2D User s Manual no content intended for this page...
Page 74: ...Chapter 6 Specifications 66 TMO2D User s Manual no content intended for this page...
Page 87: ...Appendix A Outline and Installation Drawings TMO2D User s Manual 79...
Page 89: ...Appendix B Menu Maps 81 TMO2D User s Manual no content intended for this page...
Page 94: ...Appendix B Menu Maps TMO2D Digital Display User s Manual 86 Figure 30 Factory Setup Menu...
Page 95: ...Appendix B Menu Maps TMO2D Digital Display User s Manual 87...
Page 99: ...76 TMO2D User s Manual no content intended for this page...
Page 100: ......