TMO2D User’s Manual
11
Chapter 2. Installation
Figure 6: Wiring Connections to TB1 Block
CAUTION!
Do not make any connections to unassigned or unused terminals.
2.3.3.2
CE Mark Compliance
IMPORTANT:
CE Mark compliance is required only for units used in EU countries.
For CE Mark compliance, you must use shielded cable and connect the shield to the stand off as shown in Figure 7 on
page 12. After you make all the necessary electrical connections, seal the unused cable entry holes with standard
conduit plugs or their equivalent.
Note:
If you make the modifications as discussed here, your unit will comply with the EMC Directive 2004/108/EC.
Note:
For CE compliance, the I/O cables must be shielded. The shields are to be grounded within the TMO2D to the
closest location. Shielded cable is not required when installations include metal conduit.
2
3
4
1
+24 V/1 AMP 1
Return 2
Oxygen 3
None 4
Red
Black
White
Summary of Contents for TMO2D
Page 1: ...panametrics com Process Analyzers 910 084 D August 2021 TMO2D Digital Display User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Chapter 1 Features and Capabilities 2 TMO2D User s Manual no content intended for this page...
Page 24: ...Chapter 2 Installation 16 TMO2D User s Manual no content intended for this page...
Page 26: ...Chapter 2 Installation TMO2D User s Manual 18...
Page 64: ...Chapter 4 Basic Programming 56 TMO2D User s Manual no content intended for this page...
Page 74: ...Chapter 6 Specifications 66 TMO2D User s Manual no content intended for this page...
Page 87: ...Appendix A Outline and Installation Drawings TMO2D User s Manual 79...
Page 89: ...Appendix B Menu Maps 81 TMO2D User s Manual no content intended for this page...
Page 94: ...Appendix B Menu Maps TMO2D Digital Display User s Manual 86 Figure 30 Factory Setup Menu...
Page 95: ...Appendix B Menu Maps TMO2D Digital Display User s Manual 87...
Page 99: ...76 TMO2D User s Manual no content intended for this page...
Page 100: ......