PanaFlow™ MV80, MV82 & MV84 User’s Manual
83
Chapter 3. Operating Instructions
3.5.7
The Totalizer #1 Menu
Use the Totalizer Menu to configure and monitor the totalizer. The totalizer maximum count is 999,999,999 at which
point it will roll over to 0. The totalizer output is a 50 millisecond (.05 second) positive pulse (relay closed for 50
milliseconds). The totalizer cannot operate faster than one pulse every 100 millisecond (.1 second). A good rule to
follow is to set the unit per pulse value equal to the maximum flow in the same units per second. This will limit the
pulse to no faster than one pulse every second.
As an example of how to set an output, refer to
. This example shows how to set the totalizer to
track mass flow in kg/sec.
Note:
All outputs are disabled while using the Setup Menus.
First, set the desired units of measurement:
1.
Use
keys to move to the Units Menu (see to
2.
Press the key until Mass Flow Unit appears. Press
ENTER
.
3.
Press the key until kg appears in the numerator. Press the
key to move the underline cursor to the
denominator. Press the key until sec appears in the denominator. Press
ENTER
to select.
4.
Press the key until Units Menu appears.
Second, set the pulse output:
1.
Use
keys to move to the Totalizer Menu.
2.
Press the key until Totaling appears.
3.
Press
ENTER
and press the key to select Mass. Press
ENTER
.
4.
Press the key to set the pulse output in the units you have selected for mass flow of kg/sec. Press
ENTER
and use
keys to set the pulse value equal to the maximum flow in the same units per second. Press
ENTER
.
5.
To reset the totalizer, press the key until Reset Total? appears. Press
ENTER
and the key to reset the totalizer if
desired. Press
ENTER
.
6.
Press the
EXIT
key and answer YES to permanently save your changes.