Appendix C. Flow Meter Calculations
158
PanaFlow™ MV80, MV82 & MV84 User’s Manual
C.2
Insertion Flow Meter Calculations
C.2.1
Flowing Velocity
V
f
f
K
c
-----
=
C.2.2 Volume Flow Rate
Q
V
V
f
A
=
C.2.3 Mass Flow Rate
Q
M
V
f
A
=
Where:
A
= Cross sectional area of the pipe (ft
2
)
f
= Vortex shedding frequency (pulses/sec)
K
C
= Meter factor corrected for Reynolds Number (pulses/ft)
Q
V
= Volume flow rate (ft
3
/sec)
Q
M
= Mass flow rate (lbm/sec)
V
f
= Flowing velocity (ft/sec)
= Density (lbm/ft
3
)