Chapter 2. Installation
46
PanaFlow™ MV80, MV82 & MV84 User’s Manual
2.6.6
Remote Electronics Wiring
The remote electronics enclosure should be mounted in a convenient, easy to reach location. For hazardous location
installations, make sure to observe agency requirements for installation. Allow some slack in the interface cable
between the junction box and the remote electronics enclosure. To prevent damage to the wiring connections, do
not put stress on the wire connections at any time.
The meter is shipped with temporary strain relief glands at each end of the cable. Disconnect the cable from the
meter's terminal block inside the junction box only (see
below). Do not disconnect the cable from the
remote electronics enclosure. Remove both strain relief glands and install appropriate conduit entry glands and
conduit. After the installation is complete, re-connect each labeled wire to the corresponding terminal position on
the junction box terminal block. Make sure to connect the shield for each wire pair shield.
RED 5
BLK 5
BLK 6
RED 6
SHLD 5&6
SHIELD
T
1
T
2
T
3
T
4
E
1
S
1
S
2
E
2
TEMPERATURE
RED 3
RED 4
BLK 4
BLK 3
SHLD 3&4
SHIELD
P
1
P
2
P
3
P
4
PRESSURE
SHLD 1&2
RED 2
BLK 2
RED 1
BLK 1
PWR
1
2
SENSOR V
SENSOR V
GND
SHIELD
VORTEX
Note:
Incorrect connections will cause the meter to malfunction.
Note:
The numeric code in the junction box label matches the wire labels.
Figure 48: Flow Meter Junction Box Sensor Connections