PanaFlow™ MV80, MV82 & MV84 User’s Manual
81
Chapter 3. Operating Instructions
pressing the
keys until Alarm [1] appears. The lower line displays the mass flow rate at which the alarm
activates.
Note:
All outputs are disabled while using the Setup Menus.
First, set the desired units of measurement:
1.
Use
keys to move to the Units Menu (see
).
2.
Press the key until Mass Flow Unit appears. Press
ENTER
.
3.
Press the key until lb appears in the numerator. Press the
key to move the underline cursor to the
denominator. Press the key until hr appears in the denominator. Press
ENTER
to select.
4.
Press the key until Units Menu appears.
Second, set the alarm:
1.
Use
keys to move to the Alarms Menu.
2.
Press the key until Relay Alarm 1 appears.
3.
Press the
key to access Measure selections. Press
ENTER
and use the key to select Mass. Press
ENTER
.
4.
Press the
key to select the alarm Mode. Press
ENTER
and use key to select HIGH Alarm. Press
ENTER
.
5.
Press the
key to select the value that must be exceeded before the alarm activates. Press
ENTER
and use
keys to set 100 or 100.0. Press
ENTER
.
6.
Press the
EXIT
key to save your changes. Alarm changes are always permanently saved, and up to three relay
alarm outputs are available depending on meter configuration.