Chapter 3. Operating Instructions
80
PanaFlow™ MV80, MV82 & MV84 User’s Manual
3.5.5
The Display Menu
Use the
Display Menu
shown in
below to set the cycle time for automatic screen sequencing used in Run
Mode, change the precision of displayed values, smooth the values or enable or disable each item displayed in the
Run Mode screens.
As an example of how to change a Run Mode display item, refer to
below. This example shows how to
remove the temperature screen from the Run Mode screens.
Note:
All outputs are disabled while using the Setup Menus.
1.
Use the
keys to move to the Display Menu.
2.
Press the key until Mf Vf Pr Te De T appears.
3.
Press
ENTER
to select. Press the
key until the cursor is positioned below Te.
4.
Press the key until N appears. Press
ENTER
to select.
5.
Press
EXIT
and then
ENTER
to save changes and return to the Run Mode.
Figure 99: The Display Menu
3.5.6
The Alarms Menu
As an example of how to set an output, refer to
. This example shows how to set
Relay Alarm 1
to
activate if the mass flow rate is greater than 100 lb/hr. You can check the alarm configuration in Run Mode by