3.3.1 User roles
There are four different types of user with different access
levels as defined below
•
Administrator:
Administrators have full permissions and
can access and modify all pages of the web console.
Administrators can create new accounts on the system,
including other administrators.
•
Field Service:
Field service accounts have the highest
level of access with the added ability to update
software via the Features page and upload/download
configuration files for meter validation on the service
page. Panametrics service personnel will typically create
a field service account so they can access the system
without requiring an administrator to be present. They will,
however, not be able to add, delete or modify other user
accounts set up on the system by an administrator.
•
Operator:
Operators can access and modify the
monitor, flare, and fuel pages, so that they may
observe and change the flare.IQ operating parameters
as required. Operators do not have access to the
network configuration page. An operator may
change their own password on the accounts page.
•
Viewer:
Viewers have read-only access to the
monitor page and may change their own password
on the accounts page.
Figure 14 shows how new users can be added by creating
new accounts for them along with the username and
password requirements. During the initial configuration,
the Panametrics field service engineer will log in using their
credentials but can help the administrator set up other
accounts as required.
3.3.2 Network setup
There are two ethernet switches on the top flare.IQ
module, labeled configure/monitor (ethernet 1) and
LAN B flare control/DCS (ethernet 2). The web console
uses the LAN A (ethernet 1) jack on each DPU. The Modbus
TCP connection to the DCS is only supported on the LAN B
(ethernet 2) jack. DHCP is disabled. By default, the ports
are at the following addresses:
• LAN A: IP address: 192.168.100.100,
netmask 255.255.255.0, gateway = 192.168.100.1
• LAN B: IP address: 192.168.100.200,
netmask 255.255.255.0, gateway = 192.168.100.1
Table 2: Factory set default IP addresses
DPU
module #
DPU config
IP address
Modbus
IP address
1
192.168.100.100
192.168.100.200
2
192.168.100.101
192.168.100.201
3
192.168.100.102
192.168.100.202
4
192.168.100.103
192.168.100.203
5
192.168.100.104
192.168.100.204
6
192.168.100.105
192.168.100.205
Figure 14: Web console accounts page showing procedure of setting up new accounts with
different access levels.
19
Summary of Contents for flare.IQ Generation 2
Page 1: ...flare IQ Generation 2 Operation and maintenance manual...
Page 2: ...2...
Page 3: ...flare IQ Generation 2 Operation and maintenance manual 910 350 rev A June 2020 3...
Page 4: ...no content intended for this page 4...
Page 38: ...Figure 30 System wiring diagram 38...
Page 48: ...48...