Page 56 of 92 PAC L.P., All Rights Reserved, 2015
R6000S Operation Manual: Rev. G
2. Make sure the Sulfur Level Meter has a display. If Auto Zeroing didn't occur with power up, press
Auto Zero button and observe that the Zeroing Light illuminates and that the number on the Level
Meter is adjusted towards zero (0.0).
3. Set the Sensitivity knob to 1.
4. Set the High/Low Gain switch to High.
5. Turn ON the PMT High-Voltage.
6. Switch ON power to the UV lamp.
7. Wait approximately 30 minutes to allow the UV lamp and PMT to stabilize.
UV LIGHT RADIATION HAZARD!
MINOR OR MODERATE INJURY COULD OCCUR IF NOT AVOIDED!
A UV light radiation hazard exists.
HOW TO PROCEED: Avoid hazard by NOT LOOKING DIRECTLY INTO ANY UV
LIGHT SOURCE. Wear appropriate UV eye protection.
8. Depress the Auto Zero button to bring the Level Meter to zero.
9. Loosen the UV lamp housing "null adjustment screw" on the housing bracket.
10. Slowly move the UV lamp housing back and forth along the slide bracket while observing the signal
level meter.
11. Locate the position that produces the highest (most positive) reading on the level meter. The UV lamp
is now at the most optically sensitive position.
12. Tighten the UV lamp housing "null adjustment screw" on the housing bracket.
13. Replace the top cover of the detector module.
14. Perform the startup procedure to confirm proper function of the lamp.
15. Recalibrate the instrument.
16. Resume normal operation.
4.2.6
UV Lamp Replacement
In the event the sulfur UV lamp needs replacement, the following steps should be taken.
1. Turn the power switch to the OFF position and remove the power cord from the detector module,
following page.
Summary of Contents for R6000S
Page 1: ...R6000S Operation Manual Revision G www paclp com...
Page 4: ...Page 2 of 92 PAC L P All Rights Reserved 2015 R6000S Operation Manual Rev G LEFT BLANK...
Page 6: ...Page 4 of 92 PAC L P All Rights Reserved 2015 R6000S Operation Manual Rev G LEFT BLANK...
Page 12: ...Page 10 of 92 PAC L P All Rights Reserved 2015 R6000S Operation Manual Rev G LEFT BLANK...
Page 44: ...Page 42 of 92 PAC L P All Rights Reserved 2015 R6000S Operation Manual Rev G LEFT BLANK...
Page 50: ...Page 48 of 92 PAC L P All Rights Reserved 2015 R6000S Operation Manual Rev G LEFT BLANK...
Page 80: ...Page 78 of 92 PAC L P All Rights Reserved 2015 R6000S Operation Manual Rev G LEFT BLANK...
Page 81: ...PAGE 79 OF 92 INTERCONNECT DIAGRAM R6000S REF B42175 R3 TP 1 1...
Page 82: ...PAGE 80 OF 92 FLOW DIAGRAM INTERCONNECT DET FURNACE REF B42134 R1 TP 1 1...
Page 86: ...MK DA QM SR PAGE 84 OF 92 ASSY SULFUR DETECTOR W COOLER REF B42025 N TP 1 1...
Page 87: ...LL DGA QM SR PAGE 85 OF 92 WIRING DIAGRAM SULFUR DETECTOR 110 120VAC REF B42146 F TP 1 1...
Page 88: ...LL QM JK PAGE 86 OF 92 WIRING DIAGRAM SULFUR DETECTOR 220 230VAC REF B42145 E TP 1 1...
Page 91: ...PDalbello JBE PAGE 89 OF 92 COMPONENT IDENT REAR PANEL FURNACE SECTION REF B42162 R3 TP 1 1...
Page 92: ...PAGE 90 OF 92 FLOW DIAGRAM FURNACE SECTION WITH OPTIONAL FURNACE BYPASS REF B42163 R1 TP 1 1...
Page 94: ...PDalbello JBE PAGE 92 OF 92 WIRING DIAGRAM FURNACE SECTION 220 230VAC REF B42144 R3 TP 1 1...