R6000S Operation Manual: Rev. G
PAC L.P., All Rights Reserved, 2015
Page 53 of 92
4. Carefully extract the PMT from the housing and unplug it from the socket.
5. Install the new PMT into the socket and insert the end plate assembly into the PMT housing.
6. Screw the preamplifier into the PMT housing.
7. Perform the removal steps in reverse order (4-1) to continue reinstalling the PMT replacement.
8. Perform the start-up procedure to confirm proper operation of the new PMT.
9. Recalibrate the instrument.
10. Resume normal operation.
4.2.3
Pre-amplifier Assembly Removal and Replacement
In the event the sulfur or nitrogen detector pre-amplifier needs replacement, the following steps should be
taken.
1. Turn the power switch to the OFF position and remove the power cord from the detector module.
ELECTROCUTION HAZARD!
DEATH OR SERIOUS INJURY WILL OCCUR IF NOT AVOIDED!
An electrocution hazard exists because severe electrical shock hazards exist
when working around the high-voltage and/or 115/230 VAC components while
power is applied.
HOW TO PROCEED: Avoid hazard by turning off the unit with the ON/OFF
switch and remove the power connection or power cord from the unit to the
power source. Also, follow any required lock-out/tag-out procedures your
company policy requires.
To prevent electrical shock hazard DO NOT operate this equipment in rain or
wet conditions.
2. Use the Sulfur Detector Removal and Replacement procedure, previous section, this chapter, to
remove the detector.
3. Unscrew the preamplifier assembly from the PMT housing.
1. The PMT is extremely light sensitive. All operations involving exposure of the PMT should be conducted in
darkness or in a very low light environment.
2. Make sure that all interconnections are properly installed. Improper connections may cause damage to system
components.
Summary of Contents for R6000S
Page 1: ...R6000S Operation Manual Revision G www paclp com...
Page 4: ...Page 2 of 92 PAC L P All Rights Reserved 2015 R6000S Operation Manual Rev G LEFT BLANK...
Page 6: ...Page 4 of 92 PAC L P All Rights Reserved 2015 R6000S Operation Manual Rev G LEFT BLANK...
Page 12: ...Page 10 of 92 PAC L P All Rights Reserved 2015 R6000S Operation Manual Rev G LEFT BLANK...
Page 44: ...Page 42 of 92 PAC L P All Rights Reserved 2015 R6000S Operation Manual Rev G LEFT BLANK...
Page 50: ...Page 48 of 92 PAC L P All Rights Reserved 2015 R6000S Operation Manual Rev G LEFT BLANK...
Page 80: ...Page 78 of 92 PAC L P All Rights Reserved 2015 R6000S Operation Manual Rev G LEFT BLANK...
Page 81: ...PAGE 79 OF 92 INTERCONNECT DIAGRAM R6000S REF B42175 R3 TP 1 1...
Page 82: ...PAGE 80 OF 92 FLOW DIAGRAM INTERCONNECT DET FURNACE REF B42134 R1 TP 1 1...
Page 86: ...MK DA QM SR PAGE 84 OF 92 ASSY SULFUR DETECTOR W COOLER REF B42025 N TP 1 1...
Page 87: ...LL DGA QM SR PAGE 85 OF 92 WIRING DIAGRAM SULFUR DETECTOR 110 120VAC REF B42146 F TP 1 1...
Page 88: ...LL QM JK PAGE 86 OF 92 WIRING DIAGRAM SULFUR DETECTOR 220 230VAC REF B42145 E TP 1 1...
Page 91: ...PDalbello JBE PAGE 89 OF 92 COMPONENT IDENT REAR PANEL FURNACE SECTION REF B42162 R3 TP 1 1...
Page 92: ...PAGE 90 OF 92 FLOW DIAGRAM FURNACE SECTION WITH OPTIONAL FURNACE BYPASS REF B42163 R1 TP 1 1...
Page 94: ...PDalbello JBE PAGE 92 OF 92 WIRING DIAGRAM FURNACE SECTION 220 230VAC REF B42144 R3 TP 1 1...