Handbook
November 2011
/
59-UMC0071
/
Issue
01
Mercury
iTC
Page 93
BASIC OPERATING INSTRUCTIONS
Original Instructions
6.7
Changing control-loop parameters
6.7.1
Entering PID parameters manually
1
Tap
Control
on the Home page.
2
Tap the
PID
field until it displays
Manual
.
3
Tap the
P
field and enter a new proportional band value (in K), if required. Enter 0
for On/Off control.
4
Tap the
I
field and enter a new integral value (in minutes), if required. Enter 0 to turn
off the integral action.
5
Tap the
D
field and enter a new derivative value (in minutes), if required. Enter 0 to
turn off the derivative action.
6
Tap
Home
to return to the Home page.
6.7.2
Creating a PID table
.