THEORY OF CONTROL LOOPS
Original Instructions
12.2.7
PID tables
Two alternative methods can be used to provide PID parameters to a control loop in the
Mercury
iTC cryogenic environment controller:
Manual PID
The PID parameters are entered directly on the
Control Loop Configuration
page
). The same parameters are used for the entire temperature range
over which the control loop operates.
Auto PID
The PID parameters are obtained from a PID table (see
). This method
allows the PID parameters to vary with the controlled temperature.
12.2.7.1
PID table operation
shows an example PID table containing two rows. The values shown in this
table are not from a real system.
Figure 12-5 PID table operation
12.2.7.2
Entering a PID table
1
Tap the panel on the Home screen that is to be configured. The
Control Loop
Configuration
page is displayed.
2
Tap the
PID
field and set it to
Auto
.
3
Tap
PID Table
. The
PID Table Editor
page is displayed (see
4
If a PID table is displayed, tap
Save
to save the present table (see
5
Select the last row in the table and tap the
-
button. Tap
OK
in the confirmation field
that appears to delete the row.
6
Continue deleting rows until the table is empty.
7
Tap
+
to insert a blank row.
8
Tap the first cell in the table and enter the lowest temperature to be controlled. The
next cell in this row is displayed.