USER INTERFACE
Original Instructions
The page contains the following fields and controls:
Sensor
Select a sensor from a drop-down menu. Selecting a sensor enables other controls
on the page.
Figure 3-22 Control Loop Configuration page with a sensor selected
Heat (%)
Select
Manual
or
Automatic
control of the heater. If
Manual
is selected, set a
percentage output power for the heater in the numeric field.
Flow (%)
Select
Manual
or
Automatic
control of the gas-flow controller. If
Manual
is selected,
set a percentage flow in the numeric field.
Set (K)
If the
Heat (%)
field is set to
Auto
, set a temperature setpoint (in Kelvin) in this field.
PID
Select either
Manual
or
Automatic
PID values (see
for information on PID
control loops).
P
If
PID
is set to
Manual
, enter a P parameter value (in K) in this field.
If
PID
is set to
Automatic
, this field displays the P parameter value that has been
obtained from the PID table.
I
If
PID
is set to
Manual
, enter an I parameter value (in minutes) in this field.
If
PID
is set to
Automatic
, this field displays the I parameter value that has been
obtained from the PID table.
D
If
PID
is set to
Manual
, enter a D parameter value (in minutes) in this field.
If
PID
is set to
Automatic
, this field displays the D parameter value that has been
obtained from the PID table.
Sweep
button
Tap once to program temperature sweeps (see
). This button is only
active if
Heat
is set to
Auto
.