USER INTERFACE
Original Instructions
3.2.12
General Settings tab
Figure 3-26
General Settings tab
The page contains the following fields and controls:
Home Screen Revert Option
If Home Screen Revert mode is set
On
, the interface reverts to the Home page after
a wait period, if no user activity is detected. If this field is set
Off
, the present page is
displayed until the user navigates to another page.
Remote Access
Select
Local
from the drop-down menu to allow the user to control the iTC from the
GUI. Select an interface from the drop-down menu to enable remote control via the
selected interface.
SCPI/Legacy
field
This field is only displayed if an interface is selected in
Remote Access
.
Select
SCPI
to use a command set based on SCPI (Signal Orientated Measurement
Commands).
Select
Legacy
to use a command set that is compatible with legacy instruments.
The page contains the following buttons:
Alarm