INSTALLATION AND SETTING TO WORK
KMW3000C
Version 1
USER MANUAL
KMW3000CChiller v1.1.doc
Page
12
Figure 5.2.1 Electrical and Water Connections on the Front of the KMW3000C
5.2.2.1 Equipment Required
30 litres of clean, filtered tap water
Mains power cable for KMW3000C
Water pipes (x4, thickness 10 x 3 mm)
Screwdriver
5.2.2.2 Setting up Procedures
Caution
Ensure that the water connections are made such that the water flows through the X-ray
generator and tube in the correct direction. Failure to do this could cause the X-ray
generator/tube to overheat.
1. Connect the water pipes between the KMW3000C and the equipment to be cooled. Refer to
figure 5.2.2.
Figure 5.2.2 Water Pipe Connections
Closed circuit
valves
Tap water circuit valves
Modulated valve knob
Inlet
Outlet
Inlet
Outlet
Filter
X-ray Tube
X-ray generator
Outlet
Inlet
Inlet always at top
H/x Electronics H/x Transformer