SPECIFICATIONS
KMW3000C
Version 1
USER MANUAL
KMW3000CChiller v1.1.doc
Page
5
3. Specifications
3.1 Environment
Ambient temperature
10
°
C to 35
°
C
Storage temperature
10
°
C to 40
°
C
Relative humidity
10 – 95% non – condensing
Location
Inside a building
3.2 Performance Data
Cooling power
3000 W (at 12 ºC tap water inlet temperature)
Water flow rate
4.2 l/min
Pressure
0.3 - 0.8 bar adjustable
Temperature stability
±
1 ºC
Filter type
10 µm fibber filter
3.3 Electrical Services
Power connection
1/N AC 230V ± 10%, 50/60 Hz
Maximum power consumption
300 W
Maximum mains current
1.2 A
Main fuse
6.3 A
3.4 Water Supply
Reservoir water
Optically clear, neutral pH, de-ionised water
Connections
Quick-release type designed for pipes of size 10 x 3 mm
External water supply flow rate
0.6 l/min/kW (at 12 ºC tap water inlet temperature)
External water supply pressure
0.5 - 5 bar