TECHNICAL DESCRIPTION
KMW3000C
Version 1
USER MANUAL
KMW3000CChiller v1.1.doc
Page
7
4.2.2 Closed Circuit
Water from the reservoir flows through a constant-flow pump. The constant-flow pump controls the
flow of the water from the reservoir to a suction filter and a manual filter in the front of the unit.
Water then passes through the front panel of the KMW3000C to the devices to be cooled. Water
returning from the devices passes through the front panel of the KMW3000C via a manual valve.
The water then flows to the pressure sensor, the heat exchanger and back into the water reservoir.
Figure 4.2.1 Schematic of the KMW3000C chiller
Figure 4.2.2 View from above of the KMW3000C with the top cover removed
T
Status
lights
Enable
230VAC, 50Hz
24VAC, 50Hz
Control
Tap water inlet
Tap water outlet
Thermometer
Inlet
Outlet
Pressure sensor
Water flow rate sensor
Water
level
sensor
Water reservoir
Tap water filter
Temperature modulated valve
Temperature switch turns off at 38 ºC