4.Advanced User Guidebook
45
Windows Trigger menu
list:
MENU
SETTING
INSTRUCTION
Single
Mode
Windows
Set vertical channel trigger type as Windows
trigger.
Source
CH1
CH2
CH3
CH4
Select CH1 as the trigger source
Select CH2 as the trigger source
Select CH3 as the trigger source
Select CH4 as the trigger source
Threshold
Up Level
Low Level
Adjust the
M
knob or click
to set the up level
threshold.
Adjust the
M
knob or click
to set the low level
threshold.
Condition
Polarity
Positive Polarity, the oscilloscope triggers on the
positive Windows pulse.
Negative Polarity, the oscilloscope triggers on the
negative Windows pulse.
Enter: Triggers when the trigger signal enters the
specified trigger level range.
Exit: Triggers when the trigger signal exits the
specified trigger level range.
Time: Specify the hold time of the input signal after
entering
the
specified
trigger
level.
The
oscilloscope triggers when the accumulated hold
time is greater than the windows time. Available
range is 30ns-10s, default 100ns.
Mode
Holdoff
Auto
Normal
Single
Holdoff
Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred
When trigger occurs, acquire one waveform then
stop
100 ns - 10 s, adjust
M
knob or click
to set
time interval before another trigger occur, press
panel button or click
to move
cursor to choose which digit to be set.