4.Advanced User Guidebook
96
Figure 4-28 Autoscale Horizontal-Vertical multi-period waveforms
Note
:
1. When entering into Autoscale function, a autoscale indicator will be
flickering on the top left of the screen.
2. In the mode of Autoscale, the oscilloscope can self-estimate Trigger Mode
(Edge, Video). At this point, the trigger menu is not available.
3. When the input signal contains the DC component, the coupling will be set
to AC, the amplitude of the input signal should be greater than 5mV, and
the frequency should be greater than 20Hz.
4. At the mode of Autoscale, DSO is always set as DC coupling with AUTO
triggering, the holdoff is set to 100ns.
5. At the mode of Autoscale, if adjust the vertical position, voltage division,
trigger level or time scale, the oscilloscope will pause the Autoscale
function. To resume Autoscale, push the
Autoset
front panel button.
6. When video triggering, the horizontal time scale is 50us.
7. While the Autoscale is working, the settings below will be made forcibly:
The DSO will switch from the wave zoom mode to the normal mode.
In the decoding, pass/fail or XY mode, when entering into Autoscale, these
modes will be turned off.
In the STOP status, when entering into Autoscale, the status will be set to
RUN.
How to Use Built-in Help
1. Push
Help
button, the catalog will display in the screen.
2. In the bottom menu, press
Prev Page
or
Next Page
to choose help topic,
or just turn the
M
knob to choose.
3. Press
OK
to view the details about the topic, or just push the M knob.
4.
Press
Return
to go directly to the directory menu while viewing the content of the