4.Advanced User Guidebook
87
Source
CH1 CH2
CH3 CH4
Select the source
Add
Add the selected measure types (shown
at the left bottom, you could only add 8
types at most)
Remove
Meas Type
(left menu)
Select the types need to be deleted.
The selected type and source are shown
in the
Remove
menu on the right.
Remove
Remove the selected measure type
Remove All
Remove all the measures
Snapsh
ot
ON
OFF
Show all the measures of the snapshot
source
Turn off the snapshot
Source
CH1
CH2
CH3
CH4
Select the snapshot source
Set
Gating
Screen
Cursor
Select "Screen" or "Cursor" to set the cur
sor range
Statistics
ON
OFF
select "ON" or "OFF" for measurement v
alue statistics
Reset
Restart statistics on statistics
Measure
Only if the waveform channel is in the ON state, the measurement can be
performed. The automatic measurement can not be performed in the
following situation: 1) On the saved waveform. 2) On Waveform Math
waveform. 3) On the Video trigger mode.
On the Scan format, period and frequency can not be measured.
Measure the period, the frequency of the CH1, following the steps below:
1. Push the
Measure
front panel button to show the Measure menu.
2. Select
Add
in the bottom menu.
3. In the left Type menu, turn the
M
knob to select
Period
.
4. In the right menu, select
CH1
in the
Source
menu item.
5. In the right menu, select
Add
. The period type is added.
6. In the left Type menu, turn the
M
knob to select
Frequency
.
7. In the right menu, select
CH1
in the
Source
menu item.
8. In the right menu, select
Add
. The frequency type is added.
The measured value will be displayed at the bottom left of the screen