10.Technical Specifications
131
Performance Characteristics
Instruction
Channel
–channel
isolation
50Hz: 100 : 1
10MHz: 40 : 1
500MHz: 20 : 1
Time delay between
channel(typical)
150ps
Horizontal
System
Sampling rate range
Four CH
0.05 Sa/s - 1 GSa/s
Dual CH* 0.05 Sa/s
– 2.5 GSa/s
Single CH 0.05 Sa/s - 5 GSa/s
Interpolation
(Sinx)/x, x
Max Record length
When four channels are turned on, the
max record length is 100M; and max
200M for two channels; max 400M for
one channel.
Scanning speed
(S/div)
500ps/div - 1000s/div,
step by 1
– 2 - 5
Sampling rate / relay
time accuracy
±2.5 ppm max (Ta = +25
℃
±5
℃
)
Interval(
△
T)
accuracy
(DC - 100MHz)
Single
:
±(1 interval time+1 ppm×0.6
ns);
Average>16
:
±(1 interval time +1 ppm×0.4
ns)
Vertical
system
Sensitivity
1 mV/div - 10 V/div
Displacement
±
1V(1mV/div)
;
±
2V(2mV/div
~
50mV/div)
;
±
20V(100mV/div
~
500mV/div)
;
±
200V(1V/div
~
5V/div)
;
±
100V(10V/div)
;
Analog bandwidth
XDS4504
XDS4502
500 MHz
XDS4354
XDS4352
350 MHz
Single bandwidth
XDS4504
XDS4502
DC to 500 MHz
XDS4354
XDS4352
DC to 350 MHz
Low Frequency
≥10 Hz (On the BNC )
Rise time (at input,
Typical)
XDS4504
XDS4502
≤0.7 ns