(7) Once the parameters are downloaded, various set values are changed and the output value from the
flowmeter is also changed. If valves are controlled by the output of the flowmeter, change the control
loop to manual control for the sake of safety, so that the control loop may not be affected by the
output of the flowmeter.
(8) If you click "OK" button and the download is started, the message box (Fig. 3-91) will be displayed. It
is possible to confirm the status of downloading by the progress bar in the center of the window.
(9) Once the download is completed, a message box (Fig. 3-92) will be displayed. Click "OK" button.
Fig. 3-92
(10) Verify the downloaded content. Disconnect the communication following "3.3 Terminating the
Connection".
Next, connect the communication following "3.2.3 Connection". Check the data is changed to the
downloaded content following "3.6 Menu: Setup (Settings)".
Fig. 3-91