3.6.3.4 Selecting Status Output Objects (Status select)
① Click "Setup (S)" from the menu, and select "Status select" from the drop-down list of "Outputs",
then the following window (Fig. 45) will be displayed.
Fig.45
② Select for respective items.
Setting item
Detail
Setting range
Overflow error
Overflow error
ON, OFF
Hardware failure
Hardware failure error
ON, OFF
Parameter alarm
Parameter alarm
ON, OFF
Saturated alarm
Output saturation alarm (Analog, Pulse)
ON, OFF
Measurement signal alarm
Measurement signal alarm
ON, OFF
Empty pipe alarm
Measurement pipe empty detection alarm
ON, OFF
Fixed output
Fixed output state (Analog, Pulse, Contact)
ON, OFF
Self diag in progress
Self-diagnosis in process
ON, OFF
NOTE: 1. When setting error and alarm as status output, click "ON".
2. Error output is valid only when "Status output func" is set to "Error Status".
3. Change the contact output when one or more items selected are in error.
4. For the detail of error, refer to "3.11.4 Error/Alarm, Status List".
③ Click "OK" after the setting is completed to display the message box (Fig. 46). Click "OK" here
to change the settings to the inputted values, and to reflect the changed setting values in the
flowmeter's output. If the flowmeter's output is used to control valves or other such parts as safety
protocol, change the control loop to manual control so that the control loop cannot be influenced
by the flowmeter's output.
Fig.46