E-025-2-E
2
③ Click "OK" after the setting is completed to display the message box (Fig. 40). Click "OK" here
to change the settings to the inputted values, and to reflect the changed setting values in the
flowmeter's output. If the flowmeter's output is used to control valves or other such parts as safety
protocol, change the control loop to manual control so that the control loop cannot be influenced
by the flowmeter's output.
Fig.40
② Enter numeric values for respective items.
Setting item
Detail
Setting range
Pulse operation
mode
Pulse output mode selection (Select from the right column.)
Frequency: Calculate with full-scale frequency and full-
scale flow.
Weight: Calculate with pulse weight.
Frequency, Weight
Freq factor
Full-scale frequency of pulse output
Valid when pulse output mode is Frequency.
(Set value display area is highlighted.)
1 to 10000
Rate factor
Full-scale flow rate of pulse output
Valid when pulse output mode is Frequency.
(Set value display area is highlighted.)
—
Low cut
Low-cut value of pulse output
Low-cut value for preventing pulse output when flow rate
is near 0%.
Valid when pulse output mode is Frequency.
(Set value display area is highlighted.)
0.0 to 100.0%
Pulse weight
Pulse weight
Valid when pulse calculation mode is Weight.
(Set value display area is highlighted.)
—
④ The message box (Fig. 41) is displayed after changing the setting values. Click "OK" to complete
the setting process.
Fig.41
⑤ The window for setting item is displayed. To exit, click "Cancel".