6.7 Assembling the housing
1) Slide the electrode housing with the connected
cables into the greased lamination ring. It has to
engage completely!
2) The circuit board and the edge of the lamination
ring must be parallel to each other.
3) The electrode housing has engaged properly. The
circuit board and the edge of the lamination ring are
parallel to each other.
6.8 Installing the elbow joint
1) Do not remove the two joint covers!
2) Screw the elbow joint into the upper arm socket.
Make sure the friction adjustment screw and lamin
ation ring thread recess are applied opposite to
each other.
3) Set the AFB dial to minimum compensation force.