CAUTION
Signs of wear and tear on the product components
Injury due to faulty control or malfunction of the product
►
In the interest of patient safety and in order to maintain operating reliability and protect the warranty, the spe
cified service intervals must be observed (see the section "Maintenance").
CAUTION
Injury due to incorrect adjustment of the product.
Loosening of the components.
►
In case of a product with a quick-disconnect wrist, position the terminal device before use so that a slight turn
cannot disconnect the terminal device from the prosthesis.
CAUTION
Incorrect donning and doffing of the prosthesis
Risk of injury due to sudden flexion of the prosthesis.
►
Switch off the prosthesis before donning or doffing.
►
The prosthesis must always be flexed for donning and doffing.
CAUTION
Unintentional unlocking of the terminal device
Risk of injury due to releasing the terminal device from the forearm (e.g. while carrying objects).
►
When using a quick-disconnect wrist unit, position the terminal device so that slight twisting does not release
the terminal device from the forearm.
CAUTION
Charging the product with damaged power supply unit/charger/charger cable
Injury due to unexpected behaviour of the product caused by insufficient charging.
►
Check the power supply unit, charger and charger cable for damage before use.
►
Replace any damaged power supply unit, charger or charger cable.
CAUTION
Risk of pinching in the joint flexion area
Injuries due to pinching of body parts.
►
Ensure that fingers and other body parts are not in this area when bending the joint.
NOTICE
Coating, gluing or painting the prosthesis
Damage or fracture due to chemical processes.
►
The prosthesis must not be coated, glued or painted.
INFORMATION
►
The electrodes are to be placed on intact skin only and with as much electrode-skin contact as possible.
►
In case of strong interference from electronic devices, the position of the electrodes should be checked and
changed if necessary.
►
If interference cannot be eliminated, please contact the Ottobock branch responsible for your country.
►
Allow the patient to rest during the adjustment of the electrodes. Muscle fatigue leads to inconsistent results,
and therefore the therapist will tend to establish electrode settings that are excessively sensitive.
5 Scope of delivery
•
1 pc. 12K110N=* DynamicArm Plus
•
1 pc. 13E219 control electronics
•
1 pc. cable lock
•
1 pc. cable guide
41
DynamicArm Plus 12K110N=*
Scope of delivery