Sequential switching
Channel A
Switching back and forth
Channel B
“LT_A Step 3” or “Switch_B” defines the two prosthetic components for sequential switching with a switch
impulse. The switch has to be pressed and released again within the configured timer value for switching to be
successful.
Switch/LT Step Timer
This function regulates the time setting. The default setting is 1.5 seconds and can be changed at any time.
The timer starts as soon as a locking or switch position has been reached. Switching takes place only if pulling on
the 4-stage control element or actuating the switch ends before the configured time runs out.
7.3.2.8 Automatic switch-back
Automatic, time-controlled switching between three selectable prosthetic com
ponents is possible with this function. The “Automatic Timer” is activated as
soon as a component is no longer being moved. Switching to the next selected
component takes place automatically after the configured time runs out.
Function of the “Automatic Timer”
The “Automatic Timer” is activated as soon as no component is being moved anymore and switches to the next
component after the configured time runs out. The default setting for the “Automatic Timer” is 5 seconds.
7.3.2.9 Connection indicator
The connection indicator shows whether there is an active Bluetooth wireless
connection between the elbow joint and computer. The number indicates the
active COM port.
7.3.2.10 Configuration of the prosthetic component
1. Deactivation/activation of prosthesis myo-control.
2. Manual control of the prosthetic component
3. Reverse (movement reversal)
4. Adjusting the working range